Chapter 6 | DynaDrive S.U.B. specifications
Table 6.6. 3,000 and 5,000 L DynaDrive S.U.B. specifications (continued).
3,000 L 5,000 L
Support container
Overall width 228 cm (90 in.)
Overall length 164 cm (65 in.)
Overall height 345 cm (136 in.) 486 cm (192 in.)
Dry skid weight (mass) 2,525 kg (5,566 lb) 3,224 kg (7,109 lb)
Wet skid weight—rated working volume (mass) 5,518 kg (12,166 lb) 8,124 kg (18,109 lb)
General
Electrical power supply requirement
(voltage, phase, amp)
240 VAC, single, 20 A
pH & DO probe—autoclavable and single-use options
available (single-use are integrated into the BPC)
12 mm diameter x 215–235 mm insertion length x 13.5
PG (pipe) thread
Noise level < 70 dB at 1.5 m
Recommended operating parameters
Room operating temperature range 17°C to 40°C ± 0.5°C (104°F ± 0.9°F)
Motor speed 0–90 rpm
Volume range 250–3,000 L 250–5,000 L
Maximum bag pressure 34 mbar (0.5 psi)
Temperature control
Cold plate area of both plates/each plate 5,690 cm² (882 in.²)/2,845 cm² (441 in.²)
Heater plate area of both plates/each plate 1,523 cm² (236 in.²)/716 cm² (118 in.²)
Condenser TCU size Minimum 2,000 W
Temperature set point 5°C
Pressure set point 30–70 psi*
Flow rate Minimum 3 GPM
Recommended heater set point 50°C
*Assumes 1 in. diameter hose on TCU.
Thermo Scientific 69DynaDrive S.U.B. Service Guide