Appendix C Preventive Maintenance
C.1 Safety Testing
PAT testing in accordance with local regulations, should be performed on a regular
basis, (typically annually for an instrument in daily use).
C.2 Cleaning
Shock Warning
The equipment contains no user servicable parts. There is a risk of severe
electrical shock if the equipment is operated with the covers removed.
Only personnel authorized by Thorlabs Ltd and trained in the
maintenance of this equipment should remove its covers or attempt any
repairs or adjustments. Maintenance is limited to safety testing and
cleaning as described in the following sections.
Caution
The instrument contains a power supply filter. Insulation testing of the
power supply connector should be performed using a DC voltage.
Warning
Disconnect the power supply before cleaning the unit.
Never allow water to get inside the case.
Do not saturate the unit.
Do not use any type of abrasive pad, scouring powder or solvent, e.g.
alcohol or benzene.