Software
70
TOX_Manual_Process-monitoring-unit_CEP400T_en
Force sensor without DMS subprint card
1
2
4
5
6
7
8
3
9
Button, input/control panel Function
1 Active Activatingx or deactivatingo the selected
channel.
Deactivated channels are not evaluated
and not displayed in the measurement
menu.
2 Nominal Force The nominal force of the force transducer
corresponds to the force at a maximum
measuring signal.
3 Nominal force, unit Unit of nominal force (maximum of 4
characters)
4 Offset Offset value of the measuring signal for
adjusting a possible zero point offset of
the analog measuring signal of the sen-
sor.
5 Offset limit Maximum tolerated force sensor offset.
6 Forced offset NO: The process monitoring system
ready to measure directly after being
switched on.
YES: The process monitoring system
carries out an offset adjustment for the
respective channel automatically after ev-
ery start.
7 Filter Limit frequency of the measurement
channel
8 Calibrating The force sensor calibration menu opens.
9 Offset adjustment Read in the current measuring signal as
offset of the force sensor.