Setting the force sensor filter
By setting a filter value the higher frequency deviations of the measuring
signal can be filtered out.
ü The user is logged in with a suitable user level. The necessary write
permissions are available.
ü The ''Configuration -> Force sensor configuration '' menu is opened.
è Tap on the Filter input field.
w Each tap changes the value between OFF, 5, 10, 20, 50, 100, 200,
500, 1000.
Force sensor calibration
In menu ''Enter Configuration -> Configuration of force sensor\Nominal
force'' the measured electrical signal is converted to the corresponding
physical unit with the values of nominal force and offset. If the values for
nominal force and offset are not known, they can be determined via the cal-
ibration. For this a 2-point calibration is carried out. The first point here can
be the opened press with 0 kN force applied for example. The second
point, for example, can be the closed press when 2 kN force is applied. The
applied forces must be known for carrying out the calibration, for example,
which can be read on a reference sensor.
è Open the ''Enter Configuration -> Force sensor configuration\Nominal
force'' by tapping the button in ''Configuration\Configuration of
force sensor\''.