9.3 Automatic Datalogging
The 90-FL can automatically log records into the Logger. There are three automatic datalogging
modes to choose from…
1. Rate Per Day
Logs from 1 to 288 readings per day, evenly spaced throughout each 24 hour period.
Unit is dormant between readings and “wakes up” when a reading is due.
Dissolved Oxygen stirrer is switched on for 40 seconds before each reading is logged (if
Dissolved Oxygen stirrer output is enabled).
Unit continues to log until automatic datalogging is disabled, or until the memory is full.
2. Time of Day
Logs at up to12 discrete times of the day, which can be unevenly spaced throughout each 24
hour period.
Unit is dormant between readings and “wakes up” when a reading is due.
Dissolved Oxygen stirrer is switched on for 40 seconds before each reading is logged (if
Dissolved Oxygen stirrer output is enabled).
Unit continues to log until automatic datalogging is disabled, or until the memory is full.
3. Sampling Period and Duration
Logs a reading every 1 to 300 seconds for a duration of 1 to 720 minutes.
Duration can be set to log continuously until the memory is full.
Unit is turned on continuously in this logging mode.
Dissolved Oxygen stirrer is switched on continuously in this logging mode (if Dissolved
Oxygen stirrer output is enabled).
The automatic datalogging parameters of the 90-FL must first be programmed, then logging can be
started and stopped as required.