EasyManuals Logo

Trace2O HT1000 User Manual

Trace2O HT1000
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
HT1000_0.5 09/2015
47
3.5.2
General
Finding
Possible Causes
Elimination
Test result deviates from
the expected.
Chemical species not as
required.
Press arrow keys to select the
required chemical species.
No differentiation:
e.g. for the Chlorine test
there is no selection
between differentiated,
free or total.
Profi-Mode is switched on.
Switch Profi-Mode off with
Mode 50.
The pre-programmed
countdown is not
displayed.
Countdown is not activated
and/or the Profi-Mode is
activated.
Switch the countdown on with
Mode 13 and/or switch the
Profi-Mode off with Mode 50.
It seems that a method
is not available.
Method is not activated in
the user method list.
Activate the required method
in the user method list with
Mode 60.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Trace2O HT1000 and is the answer not in the manual?

Trace2O HT1000 Specifications

General IconGeneral
BrandTrace2O
ModelHT1000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals