HT1000_0.5 09/2015
47
3.5.2
General
Finding
Possible Causes
Elimination
Test result deviates from
the expected.
Chemical species not as
required.
Press arrow keys to select the
required chemical species.
No differentiation:
e.g. for the Chlorine test
there is no selection
between differentiated,
free or total.
Profi-Mode is switched on.
Switch Profi-Mode off with
Mode 50.
The pre-programmed
countdown is not
displayed.
Countdown is not activated
and/or the Profi-Mode is
activated.
Switch the countdown on with
Mode 13 and/or switch the
Profi-Mode off with Mode 50.
It seems that a method
is not available.
Method is not activated in
the user method list.
Activate the required method
in the user method list with
Mode 60.