To access the Drift Test function, press 〈.〉〈0〉 at the Setup menu.
The gauge displays the Drift Test menu:
To select a menu option, press the corresponding numeric key
or press
〈ESC〉 to return to the Setup menu.
Taking a Drift Test
To take a new drift test, press 〈1〉 at the Drift Test menu. Place
the gauge on a level surface with the source rod in the SAFE
(shielded) position, and then press
〈ENTER/START〉.
NOTE
If it has been less than three hours or more than eight
hours since the last stat test, the gauge displays an
error message. Press 〈ENTER/START〉 to return to the
Setup menu.
The gauge displays the progress of the drift test (five 4-minute
counts). Upon completion of the counts, the gauge displays the
results of the density drift test, including an indication of
whether the test passed or failed:
Press 〈ENTER/START〉 to view the results of the moisture drift
test. Follow the prompts to scroll through the results.
1. Take Drift Test
2. Review Drift Test
3. Print Drift Test
<ESC> to Exit
Density Drift Test
Avg. Counts: ####
R: #.### (Pass)
<ENTER> for Moist.