B-4 Model 3772/3771 Condensation Particle Counter
RLP Reads the laser current in milliamps
Returns: An integer from 0 to 150
Example: 70
RLL Reads the liquid level
Returns: FULL or NOTFULL and the corresponding ADC
reading. The ADC reading is an integer from 0 to
4095
Example: FULL (2471)
RMN Read the model number
Returns: 3771, 3772, 3775, 3776, 3790 or 100
Example: 3772
R0 Legacy command to read the liquid level
Returns: FULL or NOTFULL
R1 Legacy command to read the condenser temperature in
degrees Celsius
R2 Legacy command to read the saturator temperature in
degrees Celsius
R3 Legacy command to read the optics temperature in degrees
Celsius
R5 Legacy command to read the instrument status.
Returns: READY or NOTREADY
RD Legacy command to read the concentration in p/cc, 1 sec.
avg.
RV Read the version string.
Returns: Model 377x Ver B.B.B S/N AAAAAAAA
RCOUNT1 Reads the 1 second corrected counts buffer
Example: 258