Firmware Commands B-7
SCOM Setup auxiliary comport
Params: Port => 1,2,3
Baud =>
2400,4800,9600,14400,19200,28800,38400,576
00,115200
Bits => 5,6,7,8
Parity => E, O, N
Stop => 1, 1.5, 2
Example: SCOM,2,9600,7,E,1 (Set 2nd serial port to 9600,
7 bits, Even Parity, 1 Stop bit
SHOUR Set the Real Time Clock Hours (24 hour mode)
Params: hour => 0–23
Example: SHOUR,13 (sets the hour to 13)
SMINUTE Set the Real Time Clock Minutes
Params: min => 0–59
Example: SMINUTE,45 (sets minutes to 45)
SSECOND Set the Real Time Clock Seconds
Params: sec => 0–59
Example: SSECOND,0 (sets seconds to zero)
SYEAR Set the Real Time Clock Year
Params: year => 0–99
Example: SYEAR,6 (sets the year to 2006)
SDAY Set the Real Time Clock Day of the Month
Params: day => 1–31
Example: SDAY,23 (sets the day to the 23rd of the month)
SMONTH Set the Real Time Clock current Month
Params: month => 1–12
Example: SMONTH,2 (sets the month to February)
SFILL Turn on/off auto fill
Params: 0 => Off
1 => On
Example: SFILL,1 (turns on auto fill)
SDRAIN Turn drain on/off for Models 3771, 3772, 3790, CPC100
Params: 0 => Off
1 => On
Example: SDRAIN,1 (turns drain on)