Caution!
In order to avoid any injury by electrical hazard, it is
recommended that a ground fault protection device (GFCI)
be installed in the electrical panel feeding the autoclave
(local codes may make this mandatory).
Note:
The electrical network must comply with local rules or regulations.
Verify that there is an easy access to the main power switch and to the current
leakage safety relay (GFCI). The voltage supplied to the device must comply
with the label ± 5%.
3.5. Environmental Emission Information
The peak sound level generated by the autoclave is 65dBa with
background noise of 48 dBa.
The total heat per hour transmitted by the autoclave is <200Wh.
3.6. Construction
The main parts of the autoclave are made of materials as indicated
below:
Chamber is built of stainless steel 316 L.
Door is made of stainless steel 316.