PRINTER OUTPUT DESCRIPTION
Autoclave:1 Number of autoclave.
Operator :___________ To be filled in manually by operator.
11/02/2001 12:22:53 Date and time sterilization cycle ended.
Load number: 00004 Load number. Useful to determine when to clean the
chamber.
Cycle ended
D53:53 099.7°C 096.5k The time, temperature and pressure during drying.
-----------------------
E48:52 100.0°C 099.1k The time, temperature and pressure during exhaust.
-----------------------
S31:32 121.3°C 208.1k The time, temperature and pressure during sterilization.
*
* Prints sterilization data every 1 minute.
*
S12:32 121.8°C 221.1k The time, temperature and pressure during
sterilization.
S11:32 121.2°C 222.3k The time, temperature and pressure during
sterilization.
----------------------
H10:33 119.9°C 216.0k The time, temperature and pressure during heating.
*
H00:33 032.2°C 094.9k The time, temperature and pressure during heating.
----------------------
W00:00 027.8°C 094.5k The time, temperature and pressure during water inlet.
11/02/2001 11:28:49 Date and time sterilization cycle begun.
Dry time: 005min Drying time for selected program.
Ster time: 020min Sterilization time for selected program.
Ster Temp: 121°C Sterilization temperature in chamber for selected
program.
PROGRAM: 4-Liquids Selected program: 4–Liquid 121 cycle
Ver = Lab00Tn1 Number and version of the program
Legend
W
H Heating stage
S Sterilization stage
E Exhaust stage
D
C Cooling stage (only on program 6)
k kPa