Page 61 of 66 Pages
Solution
Clean the water level electrode inside the vessel.
5.2 Check levelling of the machine. (see para. 6, Installation)
6.1 Clean the water level electrode inside the vessel.
6.2 Check the levelling of the machine. (see para. 6, Installation)
7.1 Make sure the door is tightened enough. Replace the door gasket if required. (see para. 9.5
Replacing the Door Gasket)
8.1 Clean strainer according to instructions.
9.1 Make sure the paper is mounted in the right way. Only one side of the paper is printable.
(see para. 5.2,, Printer handling)
10.1 Make sure the paper is inserted in the printer.
(see para. 5.2, Printer handling)
10.2 Switch off the machine and switch it back on while pressing the feed button on the printer. If
the printer prints a test printout, the printer is O.K. and there is a problem with the electronics.
Contact your dealer to solve the problem.
If the printer does not print the test printout, there is a problem with the printer. Contact your
dealer to solve the problem.
11.1 Make sure thefeed button’ on the printer is not stuck.
Problem
5. ‘High Temp’ is displayed
6. ‘High Pres’ is displayed
7. The machine is leaking at
the door.
8. Water does not exit
chamber due to clogged
outlet strainer.
9. The printer
prints, but
nothing is
printed on
the paper.
10. The printer
does not
print.
11. When the
machine is
switched on,
the printer
feeds paper
continuously.
If the
autoclave
is equipped
with a
printer