The AIR filter is located under the mounted under the cover.
1. Unscrew the filter.
2. Disconnect the pipe.
3. Connect the new filter.
4. Screw in the new filter.
5. Ensure the new filter is all the way in and seated properly.
Note: Make sure that the arrow on the filter body points inwards, toward the
chamber. Make sure that you don’t bend the filter pipe when reattaching it.
Note: Replace the air filter every 6 months or after 1000 cycles (whichever is
the shorter period).