9.4.2 Cleaning the Drain Filter
Cautions!
Make sure the autoclave is not hot before cleaning it
Before proceeding, make sure that the electric cord
is disconnected and there is no pressure in the
chamber.
1. Clean the drain filter every month.
The DRAIN filter (1) is located inside the autoclave chamber at the bottom
far end, to reach the filter open the autoclave chamber door, remove the
tray (see below).
2. Open the autoclave chamber door, remove the tray (2), (see above).
3. Clean with a soft cloth the filter and the area around the filter.
4. Check that the autoclave function normally after the drain filter cleaning.
Note: If by cleaning the filter the result is not satisfactory, proceed to the
replacement of the drain filter (paragraph 9.4.3 below).
Page 67