EasyManuals Logo

UNI-T UTD2052CL User Manual

UNI-T UTD2052CL
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
UTD2000 Series User Manual
55
for video signals. Stable waveform display is achieved only when the correcttrigger
mode is selected.
Try changing the Couping display to High Frequency Suppression or Low
Frequency Suppression or to filter any high or low frequency noise that is interfering
with triggering.
(5) No display after pressing RUN/STOP :
Check if the trigger mode is set to NORMAL or SINGLE in the menu, and
whether the level exceeds the waveform range.
If so, move the trigger level to the centre or set trigger mode to AUTO
Press AUTO to complete the automatic setup.

Table of Contents

Other manuals for UNI-T UTD2052CL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UTD2052CL and is the answer not in the manual?

UNI-T UTD2052CL Specifications

General IconGeneral
BrandUNI-T
ModelUTD2052CL
CategoryTest Equipment
LanguageEnglish

Related product manuals