Chapter 8 - Automatic Measurement 
UTD2000CEX-II series digital fluorescence oscilloscope can automatically 
measure up to 34 parameters. Press the MEASURE button to enter the 
automatic measurement menu.
Automatic Measurement menu
Descriptions
Close all Parameters
Pop-up a frame with all 
parameters presenting 
Open/Close menu for user defined 
parameters. When it is on, define 
all required parameters with using 
the Multipurpose knob. 
Real-time dynamically indicates 
measurement parameters 
Enter advanced measurement menu 
Turn off the feature
Automatic calculation of user defined 
parameters of average, max, and min. 
Only applicable when user defined 
parameters are present. 
Restart statistics
Clear all measurement 
Options
Off
On
On/Off
Off 
Parameter 1~5
Off
On 
Functions
All Parameters
User Defined
Indicator
Advanced 
measurement
Measurement 
statistics
Reset statistics
Clear
Advanced measurement menu
Descriptions
Delay measurement
Phase measurement
FRR: time between the first rising edge of 
source 1 and the first rising edge of source 2
FRF: time between the first rising edge of 
source 1 and the first falling edge of source 2
FFR: time between the first falling edge of 
source 1 and the first rising edge of source 2
FFF: time between the first falling edge of 
source 1 and the first falling edge of source 2
LRF: time between the last rising edge of 
source 1 and the last edge of source 2
LRR: time between the last rising edge of 
source 1 and the last rising edge of source 2
LFR: time between the last falling edge of 
source 1 and the last rising edge of source 2
LFF: time between the last falling edge of 
source 1 and the last falling edge of source 2
Options
Functions
Measurement 
parameters
Source 
edge
UTD2000CEX-II User Manual