Chapter 4 _________________________________________________________________ Operation 
 
VAISALA ________________________________________________________________________ 43 
Select Units 
Use the UNIT command to select metric or non-metric output units. 
Only affects data shown on the display and service port, has no effect on 
the analog and digital outputs. This command is not available on the 
digital output models (for example, GMW95). 
UNIT [x]<cr> 
Selects the unit type to output: 
m = metric units, for example, Celsius
 
n = non-metric units, for example, Fahrenheit 
 
Example: 
>unit m 
Unit                : Metric 
 
 
Analog Output Settings 
NOTE 
If you want to change analog output settings, you must set the 
transmitter’s configuration DIP switch 8 to position 
Custom. If the 
DIP, the settings are view-only. 
 
NOTE 
Commands for configuring analog outputs are not available on digital 
output models (for example, GMW95). 
 
Set Analog Output Mode 
Use the AMODE command to set the analog output mode and error 
level. Note that you cannot change between analog output types, for 
example, from voltage to current output. 
AMODE [channel lo_value hi_value error_value]<cr> 
Analog output channel, 1 ... 3. 
Low limit of the channel. 
High limit of the channel. 
Error value of the channel.