Error Messages
Table 13 Error Messages
Error Text
on Display
Error ID on
Serial Line
Possible Cause
Remedy
HTM10 01
23
Communication
problem with
HTM10 module.
Check that the HTM10
module sits firmly in place.
Remove and reconnect.
HTM10 04
21
Problem with
humidity
measurement.
- Check for missing or
damaged HUMICAP
®
sensor on the HTM10
module.
- Check for condensation
on the HUMICAP
®
sensor. Wait for the
sensor to dry out.
HTM10 03
4
Internal problem
with HTM10
module.
- Restart the transmitter.
- Replace the module if
unable to remove the
problem.
HTM10 05
22
HTM10 06
42
GM10 01
79
Communication
problem with
GM10 module.
Check that the GM10 module
sits firmly in place. Remove
and reconnect.
GM10 03
81
Internal problem
with GM10
module.
- Restart the transmitter.
- Replace the module if
unable to remove the
problem.
GM10 04
82
GM10 10
28
GM10 11
88
GM10 12
89
Dirty or damaged
GM10 module.
Replace module.
Internal 1
1
Internal problem
with the
transmitter.
- Restart the transmitter.
- Restore the factory
settings using service
port or DIP switches if
reset does not help.
Internal 2
2
Internal 3
3
Internal 4
43