User's Guide _______________________________________________________________________ 
64  ___________________________________________________________________ M211659EN-C 
Notes for CO
2
 Adjustment 
1-point CO
2 
adjustment affects either offset or gain, depending on the gas 
concentration. 1-point adjustment below 700 ppm affects the 
measurement offset, above 700 ppm it affects the gain.  
2-point adjustment will adjust both offset and gain. Choose the 
adjustment points as follows: 
-  For first point, use as low concentration as possible. Preferably 0 ppm, 
must be below 700 ppm. 
-  For second point, use as high concentration as is relevant for your 
application. Must be within 700 ... 5000 ppm. 
Using Calibration Gas vs. Ambient Gas 
You can adjust using a calibration gas with a known concentration 
(recommended), or adjust using the ambient gas: 
-  When using a calibration gas, feed the gas to the adjustment inlet at a 
flow rate of 0.3 ... 0.7 L/min. For location of the inlet, see Figure 2 on 
page 15. After connecting the gas supply to the field adjustment inlet, 
wait for at least three minutes for the reading to stabilize. 
-  When using ambient gas for adjustment, avoid breathing near the 
transmitter during the adjustment. 
Regardless of the CO
2
 reference used, it is important to let the CO
2
 
measurement stabilize before performing the adjustment. Three minutes 
is typically enough. Also allow for the stabilization period after 
adjustment, before verifying the result. 
Effect of Temperature on CO
2
 Measurement 
The device is fully temperature compensated using internal temperature 
measurement.