User's Guide _______________________________________________________________________ 
78  ___________________________________________________________________ M211659EN-C 
4.  Reconnect the transmitter cover to the mounting base so the 
transmitter powers up. Check the screen after power-up: when the 
DIP switches are in factory reset position, you will see a 
notification text. 
5.  Disconnect the transmitter cover again. 
6.  Set the DIP switches to the positions they were before. If you are 
restoring the factory configuration on a factory customized analog 
output model (GMW90 or GMW90R), leave DIP number 8 to the 
Custom position to avoid the custom configuration from being 
overwritted by the default settings at transmitter startup. For more 
information about this, see section Changing Between DIP and 
Custom Configuration on page 21. 
7.  Reconnect the transmitter cover to the mounting base. Check the 
startup screens to verify the configuration.  
Reverting to Factory Settings Using 
Service Port 
Use the FRESTORE command to restore the transmitter to factory 
settings. 
FRESTORE<cr> 
NOTE 
If you are restoring the factory configuration on a factory customized 
analog output model (GMW90 or GMW90R), leave DIP number 8 to the 
Custom position to avoid the custom configuration from being 
overwritted by the default settings at transmitter reset. For more 
information about this, see section Changing Between DIP and Custom 
Configuration on page 21. 
 
NOTE 
After using the FRESTORE command, reset the transmitter using the 
RESET command. 
 
Example: 
>pass 9000 
>frestore 
Restoring HTM10 factory parameters 
HTM10 factory parameters restored 
Restoring GM10 factory parameters 
GM10 factory parameters restored 
Restoring GMW95R factory settings 
110/110 parameters restored 
>reset 
Resetting 
GMW95R / 1.1.28.5849 / XM90 
>