Chapter 6 ____________________________________________________________ Troubleshooting
VAISALA ________________________________________________________________________ 75
Viewing Error Messages on Serial Line
View Currently Active Errors
Use the ERRS command to view currently active errors on the serial
line:
ERRS<cr>
Example:
>errs
NO ERRORS
View Error Table
Use the ERRT command to view the table of possible transmitter errors.
The table includes error ID, error count since last reset, level, current
state, and error text.
Critical errors require a transmitter reset to recover. Other errors may be
recoverable if their cause is removed.
ERRT<cr>
Example:
>errt
Id: N: Level:State: Error text
1: 0: CRITICAL:OFF: FLASH memory corrupted
2: 0: CRITICAL:OFF: Parameter read (using defaults)
3: 0: CRITICAL:OFF: Parameter write
4: 0: CRITICAL:OFF: HTM10 03 FLASH Corrupted
21: 0: ERROR:OFF: HTM10 04 RH measurement
22: 0: ERROR:OFF: HTM10 05 T measurement
23: 0: ERROR:OFF: HTM10 01 Continuous communication failure
42: 0: ERROR:OFF: HTM10 06 Device Descriptor mismatch
43: 0: CRITICAL:OFF: Factory parameter memory not consistent
79: 0: ERROR:OFF: GM10 01 Continuous communication failure
81: 0: CRITICAL:OFF: GM10 03 FLASH Corrupted
82: 0: CRITICAL:OFF: GM10 04 MEM I
28: 0: ERROR:OFF: GM10 10 Device Descriptor mismatch
88: 0: ERROR:OFF: GM10 11 Module uncalibrated
89: 0: ERROR:OFF: GM10 12 CO2 measurement

Table of Contents

Other manuals for Vaisala GMW95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala GMW95 and is the answer not in the manual?

Vaisala GMW95 Specifications

General IconGeneral
BrandVaisala
ModelGMW95
CategoryTransmitter
LanguageEnglish

Related product manuals