Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 49
Set Output Interval
Use the INTV command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the R command has
been given.
INTV [n xxx]<cr>
where
n
=
time interval, range 0 ... 9999.
xxx
=
time unit = "S", "MIN", or "H"
The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the transmitter produces them, without additional
delay.
Example:
>intv 1 min
Output interval : 1 min