3. To adjust temperature measurement:
a. If you did not adjust humidity measurement, select the Calibration tab and select
Start calibration to start the calibration mode.
b. In the T adjustment section of the page, select Restore factory adjustment to remove
any existing non-factory adjustments.
c. Insert the probe head in the reference environment for T adjustment point 1.
d. Wait for temperature measurement to stabilize fully. This may take more than 30
minutes. Monitor the reading to see when the measurement has stabilized.
e. Enter the value of the first reference into the Reference value, point 1 field. Select
outside the text box when done. The probe automatically enters the measured value
for the adjustment point.
f. If you want to adjust temperature measurement in two points, repeat steps
step 3.cstep 3.f for T adjustment point 2.
g. Select Activate adjustment to store the adjustment in the probe.
4. In the Calibration information section of the page, update the Calibration date and
Calibration text fields.
5. Select Stop calibration to end the calibration mode.
HMP Series with MMP8 and TMP1 User Guide M212022EN-F
60