5. Commissioning
5.1 Turning on DGA monitor
Verify that the DGA monitor is correctly wired and grounded before
turning it on.
CAUTION!
If the circuit breakers will not stay in the ON position, turn o Mains
power to the DGA monitor immediately and inspect the Mains and OPTPSU
power wiring. The circuit breakers may be tripping because of a loose wire or
incorrect connection.
CAUTION!
1. Verify that the circuit breakers Main and Heat inside the DGA monitor are turned on.
2. Turn on Mains power from the external disconnection device.
3. Turn on the circuit breaker F1 Main and switch S1 inside the power supply unit.
4. Check the DC OK LED on the power supply:
If the LED is lit solid green, OPTPSU power to the DGA monitor is successfully turned
on.
If the LED keeps blinking, it is likely that the 24 V DC connection to DGA monitor is
wired incorrectly. Turn o switch S1 and Mains power, and correct the problem before
attempting to power on the DGA monitor again.
5.2
 Connecting to user interface
Laptop computer with:
RJ45 Ethernet connector
Web browser (Google Chrome, Microsoft Edge, or Mozilla Firefox)
Ethernet cable with RJ45 connectors for temporary use
Administration password for this DGA Monitor
1. Connect your computer to the same network as the DGA monitor. If you are connecting
locally, connect the network cable between your computer and the port marked ETH1 on
the processing unit inside the DGA monitor.