General
2-4 ForceTriad User’s Guide
Fire/Explosion Hazard
Notice
Connect the power cord to a properly grounded power receptacle having the
correct voltage. Otherwise, product damage may result.
Important
If required by local codes, connect the energy platform to the hospital
equalization connector with an equipotential cable.
Warning
Danger: Explosion Hazard Do not use electrosurgery in the presence of
flammable anesthetics.
Fire Hazard Do not place active instruments near or in contact with flammable
materials (such as gauze or surgical drapes). Electrosurgical instruments that are
activated or hot from use can cause a fire. When not in use, place electrosurgical
instruments in a safety holster or safely away from patients, the surgical team,
and flammable materials.