Performance Characteristics
9-6 ForceTriad User’s Guide
Duty Cycle
Under maximum power settings and rated load conditions, the ForceTriad energy
platform is capable of operating a duty cycle of 25%, defined as 10 seconds active
and 30 seconds inactive, in any mode for a period of 4 hours.
Low Frequency (50/60 Hz) Leakage Current
Deactivation impedance User selectable: 1,500 , 1,800 , 2,000 or
2,200
Measurement accuracy ± 14% of Full Scale activation impedance while
keying active above 10 watts
+ 50%, -14% of full scale activation impedance
while keying active at 10 watts or below
± 5% of Full Scale activation impedance while
keying inactive
Keying delay User selectable in 500 msec increments from
0 sec to 2.5 sec
Caution
Use of duty cycles greater than 25% (10 seconds active followed by 30 seconds
inactive) will increase the risk that heat build-up under a return electrode may be
high enough to injure the patient. Do not continuously activate for longer than one
minute.
Enclosure source
current, ground open
< 300 µA
Source current, patient
leads, all outputs
Normal polarity, intact ground: < 10 µA
Normal polarity, ground open: < 50 µA
Reverse polarity, ground open: <50 µA
Mains voltage on applied part: < 50 µA
Sink current at high
line, all inputs
< 50 µA