Output Characteristics
Technical Specifications
ForceTriad User’s Guide 9-13
Output Characteristics
Maximum Output for Bipolar, Monopolar, and LigaSure Modes
Power readouts agree with actual power into rated load to within 15% or 5 watts,
whichever is greater.
* An indication of a waveform’s ability to coagulate bleeders without a cutting effect.
Mode Open Circuit Peak
Voltage (max)
Open Circuit P–P
Voltage (max)
Rated Load
(max)
Power
(max)
Crest
Factor*
Duty
Cycle
Bipolar
Low
Standard
Macro
250 V
175 V
250 V
500 V
350 V
500 V
100 Ω
100 Ω
100 Ω
95 W
95 W
95 W
1.42
1.42
1.42
N/A
N/A
N/A
Monopolar Cut
Cut
Blend
920 V
1485 V
1840 V
2970 V
300 Ω
300 Ω
300 W
200 W
1.42
2.7
N/A
50%
Valleylab (HWD) 2365 V 4730 V 300 Ω 200 W 4.3 25%
Monopolar Coag
Fulgurate
Spray
3050 V
3625 V
6100 V
7250 V
500 Ω
500 Ω
120 W
120 W
5.55
6.6
6.5%
4.6%
LigaSure 287.5 V 575 V 20 Ω 350 W 1.42 N/A