Performance Characteristics
Technical Specifications
ForceTriad User’s Guide 9-7
High Frequency (RF) Leakage Current
Input Power
Measured with leads
recommended by Valleylab
Measured directly at the
energy platform terminals
Bipolar
RF leakage
current
< 59.2 mA
rms
< 59.2 mA
rms
Monopolar
RF leakage
current
< 150 mA
rms
< 100 mA
rms
LigaSure
leakage
<132 mA
rms
< 100 mA
rms
100–120 Volt 220–240 Volt
Maximum VA at nominal line voltage:
Idle: 52 VA
Bipolar: 450 VA
Cut: 924 VA
Coag: 530 VA
Maximum VA at nominal line voltage:
Idle: 52 VA
Bipolar: 450 VA
Cut: 924 VA
Coag: 530 VA
Input mains voltage, full regulation
range: 90–132 Vac
Input mains voltage, full regulation
range: 208–264 Vac
Input mains voltage, operating range:
85–132 Vac
Input mains voltage, operating range:
170–264 Vac
Mains current (maximum):
Idle: 0.4 A
Bipolar: 2.0 A
Cut: 7.0 A
Coag: 4.0 A
LigaSure: 5.0 A
Mains current (maximum):
Idle: 0.2 A
Bipolar: 1.0 A
Cut: 3.5 A
Coag: 2.0 A
LigaSure: 2.5 A
Mains line frequency range (nominal):
50 Hz to 60 Hz
Mains line frequency range (nominal):
50 Hz to 60 Hz
Fuses (2): 5 mm x 20 mm 8A, 250 V
fast blow
Fuses (2): 5 mm x 20 mm 8A, 250 V
fast blow
Power cord: 3-prong hospital grade
connector
Power cord: 3-prong locally approved
connector