System Alarms
Troubleshooting
ForceTriad User’s Guide 7-7
System Alarms
Most system alarms require some action on your part to correct the condition;
however, some are corrected automatically. Use the following list to determine
how to correct an alarm condition.
After correcting the alarm condition, verify that the system completes the self-test
as described in the System Setup chapter.
Description or Screen Solution
Check for stuck instrument buttons and footswitches.
Check that nothing is bumping against the touchscreens.
REM Alarm Refer to the REM Alarms in this chapter.
LigaSure Check Instrument Alert Refer to Alert Situations in Chapter 6.