Return Electrodes – REM Contact Quality Monitoring System
Monopolar Function
ForceTriad User’s Guide 4-7
REM Alarm (Visual and Audible)
If the REM system senses an alarm condition, the REM indicator flashes red and
yellow, emits two beeps, and discontinues RF energy delivery. When the alarm
condition has been corrected, the indicator illuminates green. Refer to the
Troubleshooting chapter for detailed instructions on correcting REM alarms.
When a REM alarm occurs, the large red and yellow REM icon is displayed for a
few seconds and then disappears. The small green REM icon turns red.
REM Alarm
Indicator