Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
63 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
6. Click the Dark button to acquire Dark Calibration Data.
7. Click the Bright button.
8. When the HOLD THE BUTTON FOR ~3 SEC message shows up on the status window,
press the X-ray Exposure switch. Then release the button when the RELEASE THE
BUTTON appears.
9. Do step 8) repeatedly until 4 point bright calibration data are acquired.
10. Click the V button to open View16.
11. Click each point Bright Calibration data file on View16 and make sure that the Median value
meets the standard.
12. If the standard is not satisfied, set them according to the standard by adjusting kVp0 and
kVp1.
1 Point
2 Point
3 Point
4 Point