9 Diagnosis, asset management and service
VEGAPULS 69 • Foundation Fieldbus
47251-EN-160815
Code
Text message
Cause Rectication DevSpec
Diagnosis Bits
F125
Impermissible electronics
temperature
Temperature of the
electronics in the non-
speciedrange
Check ambient tem-
perature
Isolate electronics
Use instrument with
higher temperature
range
Bit 7
F260
Error in the calibration
Error in the calibration
carried out in the factory
Error in the EEPROM
Exchanging the elec-
tronics
Send instrument for
repair
Bit 8
F261
Errorintheconguration
Error during setup
False signal suppression
faulty
Error when carrying out
a reset
Repeat setup
Repeat reset
Bit 9
F264
Installation/Setup error
Adjustment not within
the vessel height/meas-
uring range
Max. measuring range
of the instrument not
sucient
Check or correct instal-
lation and/or parameter
adjustment
Use an instrument with
bigger measuring range
Bit 10
F265
Measurement function dis-
turbed
Sensor no longer carries
out a measurement
Operating voltage too
low
Check operating voltage
Carry out a reset
Disconnect operating
voltagebriey
Bit 11
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text message
Cause Rectication DevSpec Diagnosis Bits
C700
Simulation active
A simulation is active Finish simulation
Wait for the automatic
end after 60 mins.
Bit 27
The following table shows the error codes and text messages in the
status message "Outofspecication" and provides information on
causes as well as corrective measures.
Code
Text message
Cause Rectication DevSpec State Diagno-
sis Bits
S600
Impermissible electronics
temperature
Temperature of the
electronics in the non-
speciedrange
Check ambient tem-
perature
Isolate electronics
Use instrument with
higher temperature
range
Bit 23
Function check
Outofspecication

Table of Contents