Operations & Maintenance Manual: Product Specifications
09005017ETEE REV05
Table 9. Core 15 (05700404)
GENERAL SPECIFICATIONS
Classification: Electrical Class II, Applied Part BF
Line voltage:
Range: 100–240VAC, 50 and 60Hz
(If the power cord has a third prong, the third conductor is not intended to be
a protective earth connection for the medical power supply.)
DC power supply: 12V DC, 2.5A max
Ingress protection: IP54
BATTERY SPECIFICATIONS
Battery type Lithium‑ion
Battery life
Under normal operating conditions, a fully charged, new battery lasts
approximately 90 minutes.
Charging time
Charging time off‑line will take no more than 4hours from an empty battery
to a full charge.
Rated capacity 3350mAh
Nominal voltage 7.4V
Nominal weight 110g (0.24lbs)
OPERATING & STORAGE SPECIFICATIONS
Operating Conditions Shipping & Storage Conditions
Temperature: 10–35°C (50 to 95°F) 2045°C (4 to 113°F)
Relative humidity: 1095% 1095%
Atmospheric pressure: 7001060hPa 4401060hPa
COMPONENT SPECIFICATIONS
Display TFT Color, 1366 x 768 px
B
C
D
A
Monitor (A) 39.6cm (15.6in)
Height (B) 280 mm (11in)
Width (C) 415 mm (16.3in)
Depth (D) 55.2 mm (2.2in)
Weight 3.100kg (6.8lb)
For operating conditions on video scopes, please refer to the GlideScope Video Laryngoscopes
Operations & Maintenance Manual (part number 09004940) or the GlideScope BFlex
Single‑Use Bronchoscopes Operations & Maintenance Manual (part number 09004939).