20
0900‑5017‑ETEE REV‑05
PROcEDURE 3. cHARGE THE MONITOR BATTERY
Please read the Hoiatused ja ettevaatusabinõud section before performing the following task.
The GlideScope Core Monitor includes an internal lithium battery. Verathon
recommends that you charge the
battery fully prior to first use.
Under normal operating conditions, a fully charged battery lasts approximately 135minutes for Core 10,
and 90minutes for Core 15 before it needs to be recharged. For optimal battery life, ensure that the
battery is fully charged before you try to use the monitor in battery mode. You should charge the battery at
temperatures from 10 to 35°C (50 to 95°F).
An estimate of the time remaining is shown under the battery icon. This time is based on the battery draw
and may vary depending on the number of components and accessories attached. As the battery is depleted,
the battery status bar will shrink, changing color at certain levels.
Figure 6. Battery Status Icons
Red battery bar: 0% to 10% battery life remaining. Battery must be charged.
Gold battery bar: 11% to 30% battery life remaining.
White battery bar: 31% to 100% battery life remaining.
1. Connect the video monitor 12V DC poweradapter to the power cable.
2. On the back panel of the monitor, remove the power socket cap, and then connect the 12V DC power
adapter to the power socket.
3. Plug the power supply into a hospital‑grade power outlet.
4. Allow the battery to charge. Fully charging the battery may take up to 4hours.