19
Operations & Maintenance Manual: Setting Up
09005017ETEE REV05
ADJUST THE MONITOR TILT ANGLE TENSION
If the angle of the monitor is difficult to adjust, or the monitor angle tilts down on its own, the workstation’s
monitor angle tension needs to be adjusted.
Using a 4 mm hex wrench, turn the tension adjustment screw clockwise to increase the tension, or
counter‑clockwise to decrease it.
decrease
tension
Increase
tension
ADJUST THE PREMIUM WORKSTATION ARM STRENGTH
When using the Premium Workstation, depending on the combined weight of the monitor and mounted
accessories, the strength of the articulating arms lifting spring may need to be adjusted.
1. Lift the arm to its maximum height.
2. Using a 4 mm hex wrench, turn the spring adjustment screw clockwise to increase the arms lifting
strength, or counterclockwise to decrease it. The optimal adjustment will allow the arm to remain in
position without lowering or raising on its own.
Weaker Stronger