12 | Page
Version 00
© Verder Liquids B.V
Ds500, Ds500+ Operating manual
Original instructions (en)
6. TechnicalSpecications
6.1 PumpSpecications
Detail Value
Supply voltage 100-240V50/60HzAC
Power consumption 250w
Dimensions(WxHxD) 216mmx246mmx210mm
Flow range 0.1 - 500ml/min*2
Max.dischargepressure 7 bar(g)
Max.suctionpressure 2 bar (g)*
1
Max.pumpspeed 65rpm
Max.temperatureof
pumped medium
70°C
Max.dischargespeed
adjustmentrange
5000:1
Seal sealing / protects
against system pressure /
full occlusion
Yes
(up to 60 PSI static,
100 PSI dynamic)
IP rating IP66, NEMA Type 4X
Noise level <70dB(A) @1m
Weight 8.25kg
Table 5. - Pumpspecications
*
1
Inletpressureshouldnotexceed2bar(g);increasedinlet
pressurewillresultinhigherowrate.
*
2
Maxowratebecomesdynamicifowcalibrationratiois
not default 1.000
6.2 Materials
6.2.1 Product contact (cartridge)
(in the event of cartridge / tube failure)
Detail Value
Tube Verderprene
Ports PP (Polypropylene)
Connectors PP
Seals EPDM (Ethylene Propylene
Diene Monomer)
Table 6. - Materials (product contact)
6.2.2 Non-product contact (pump)
Detail Value
Pump housing 20% GF PPE/PS (Polyphenyl
Ether + Polystyrene)
Stanyl
®
(rear panel)
Drive shaft P6 (nylon) (covering SS440C
shaft)
Cartridge 20% GF PPE/PS + PA6 +
Polypropylene
Screen enclosure 20% GF PPE/PS
Screen guard Polycarbonate
Keypad Polyester
Table 7. - Materials (non-product contact)
6.3 Chemical Compatibility
See https://www.verderex.com/leadmin/les/verderex/
documents/Verderflex_Chemical_Compatibility_Guide/
Verderex_Chemical_Compatibility_Guide.pdf
Operationunderanyotherconditionswouldrequire
approvalfromthemanufacturer,otherwisethismayaect
the warranty.
6.4 Operating Conditions
6.4.1 Ambient
Detail Value
Operating temperature
range
4-45°C
Humidity 5-95%RH(non-condensing)
Max.altitude <2000m
Table 8. - Operating conditions
All units/calculations are based on operating conditions
below ≤ 2000m. For use above this height please check
with manufacturer or local representative for conrmation of
performance.