VORTEX AIR Technical Manual
Revision 6- Issued by N. Shields 24/08/2023
The pump is not receiving any power.
The fuse has blown.
The whole pump unit has failed.
Check the cable providing power to the
pump.
Check the fuse has not blown, replace as
necessary.
If the pump is receiving power and the
fuse is in tact the pump itself is faulty,
replace.
Ensure there is no debris in the pump
inlet.
Loose connection.
Incorrectly wired.
Damaged by blockage.
Check that the spade connections on the
back of the pump are properly made and
are in the correct polarity.
Remove the pump and assess for
damage to the impeller, replace if
necessary.
Check cable connection at PCB terminals.
1. The pump is not
working at all.
WATER PUMP
NOT
FUNCTIONAL
Fault
Symptoms
2. The pump is
working, but not at
full capacity.