Page | 41
VORTEX AIR Technical Manual
Revision 6- Issued by N. Shields 24/08/2023
A clip requires tightening.
A pipe has become loose or is damaged.
The diaphragm or water pump are leaking.
Remove the front panel and shine a
torch inside the machine to locate the
leak.
Replace any damaged pipework.
Tighten the appropriate clips.
Replace/repair and damaged
components).
The lid spray plumbing assembly is
compromised.
Remove the lid and check the pipework,
replace if necessary.
The inlet valve has failed. The sensor is not
working.
Check the operation of the inlet valve.
Replace if needed.
Check that the tank sensor is still in place
and is operating effectively.
The mechanical seal has failed.
The leak is caused by the motor flange
gasket, flush pipe or barbed flange fitting.
Check the flush pipe is connected to the
motor flange. Assess it for damage.
If the flush pipe is intact remove the
motor and check/replace the motor
gasket.
Check/replace the mechanical seal and
adapter assembly.
WATER LEAK
Fault
Symptoms
1. There is a leak
coming from the
internal pipework.
3. The lid assembly
is leaking.
4. The water tank is
leaking.
2. There is a leak
coming from
underneath the
hopper.