32
NOTE:
Guided Programming mode is unavailable if Day/Night programming is
enabled. In the Standard Therapy Protocol, maximum target for output
current for Normal Mode is 1.750mA. Increases above that target must be
done manually.
NOTE:
At the end of each o ce visit, the generator must be interrogated by
tapping the INTERROGATE button followed by tapping the END
SESSION button to ensure creation and accuracy of the session reports.
Standard Protocol Steps
Standard Protocol Persistent Parameter Settings
For information on creating Custom Protocols, see page 37