REMOTE OPERATION
Page 50 © Voltech 2010 PM1000+ Power Analyzer User Manual
Inrush
:MOD:INR Select in rush mode
:MOD:INR:RNG 4 Fixes the current range for in rush mode to range 4 (6.25Apk)
:MOD:INR:VRNG 4 Fixes the voltage range for in rush mode to range 4 (900Vpk)
:MOD:INR:CLR
:SEL:CLR Clear measurements
:SEL:APK+ Selects peak positive current
:SEL:APK- Selects peak negative current
Ensure equipment under test is off
:MOD:INR:CLR Clear the Apk+ and Apk-
Switch on equipment under test
:FRD? Returns Apk+ and Apk-.