REMOTE OPERATION
Page 46 © Voltech 2010 PM1000+ Power Analyzer User Manual
":SEL:REA”
Selects reactance X.
":SEL:VHM
Selects voltage harmonic series.
":SEL:AHM
Selects current harmonic series.
":SEL:HRS”
Selects integration elapsed time.
":SEL:VDC”
Selects Volts DC.
":SEL:ADC”
Selects Amps DC.
":SEL:VRNG”
Add the active voltage range to the
screen.
":SEL:ARNG"
Add the active current range to the
screen.
":SHU:INT”
Selects internal shunt.
":SHU:EXT”
Selects external shunt.
":SHU?”
Returns the current shunt setting 0 =
internal 1 = external.
":SYST:TIME?”
Returns the current time setting.
":SYST:DATE?”
Returns the current date setting.
":SYST:SET:TIME <value>”
Sets the RTC time <value> = Example 10-
10-00.
":SYST:SET:DATE <value>”
Sets the RTC date <value> = Example 12-
12-2006.
":SYST:FOR:TIME <value>”
Sets the time format <value> where 0 =
12 Hour and 1 = 24 Hour.
":SYST:FOR:DATE <value>”
Sets the RTC date format <value> = 0
mmddyyyy; 1 = ddmmyyyy; 2 = yyyymmdd
":SYST:ZERO <value>”
Set auto zero:
<value> = 0 → disabled
<value> = 1 → enabled
":SYST:ZERO?
Read auto zero state.