REMOTE OPERATION
PM1000+ Power Analyzer User Manual © Voltech 2010 Page 53
Display Data Status Enable Register (DSE)
Read by “:DSE?” and set by “:DSE <value>”.
Bit 7 – IEC_FlickIFS Enable IEC_FlickIFS bit in DSR. (Default to enabled on
power-up.)
Bit 6 – IEC_FlickBin Enable IEC_FlickBin bit in DSR. (Default to enabled on
power-up.)
Bit 5 - IEC Enable IEC bit in DSR. (Default to enabled on power-up.)
Bit 4 - OVV Enable OVV bit in DSR.
Bit 3 - OVA Enable OVA bit in DSR.
Bit 1 - NDV Enable NDV bit in DSR. (Default to enabled on power-up.)
Bit 0 - DVL Enable DVL bit in DSR. (Default to enabled on power-up.)
Standard Event Status Register (ESR)
Read by “*ESR?” or in summary by the ESB bit in STB.
Bit 5 - CME Command error; command not recognized.
Standard Event Status Enable Register (ESE)
Read by “*ESE?” and set by “*ESE <value>”. Cleared when read.
Bit 5 - CME Enable CME bit in ESR. (Default to enabled on power-up.)