2.11 Emergency shutdown of the mass spectrometer
Warning: To avoid electric shock, observe the following procedure to isolate the
instrument from the main power supply. The instrument's power switch does not isolate
it from the main power supply.
Notice: To avoid losing data, use the following procedure only in an emergency. To
reboot the mass spectrometer, follow the procedure in the section "Rebooting the mass
spectrometer".
To shut down the mass spectrometer in an emergency:
1. Press the power button on the front of the mass spectrometer.
2. Disconnect the power cable from the rear panel.