Storage clip on the overflow spur
3. Swing open the UniSpray source enclosure unit from the source mounting on the mass
spectrometer.
4. Disconnect the probe adjuster cable (yellow) from the instrument source interface panel.
5. Disconnect the impactor pin high-voltage cable from the instrument source interface panel.
6. Return the cable fittings to their storage slots on the source enclosure.
7. Carefully lift the UniSpray source off the instrument's supporting studs and store it safely.
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)