Required tools and materials
Chemical-resistant, powder-free gloves
Tray on which to place the pump
Container to catch used oil
Suitable pump oil
Warning: To avoid personal contamination with biologically hazardous or toxic compounds,
wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid burn injuries, allow the pump to cool before touching surfaces
displaying the burn warning symbol.
Warning: To avoid slip injuries, inspect the pump for oil leaks.
To add oil to the backing pump:
1. Power-off the instrument using the power button in the top, left-hand corner of the front
panel.
2. Wait approximately five minutes to allow the instrument to vent.
3. Switch off the pump and disconnect the power cable.
Warning: To avoid skeletal or muscle injury associated with lifting heavy
objects, at least two persons must lift the pump.
4. Place the pump on a tray suitable for catching dripping oil.
5. Remove the oil drain plug from the pump’s rear panel.