Simplied block diagram of RMS-Peak-FFT analysis
for the WP400 and WP400-3 probes
+
+
1 M Samples/s
for SPAN 400 kHz
Bx
FFT
Output
Real time
FFT
Axes and
total
value
Sensor ADC
Digital
lter
Square
10 i
100 k Samples/s
for SPAN 40 kHz
10 k Samples/s
for SPAN 4 kHz
1 k Samples/s
for SPAN 400 Hz
10 i
10 i
RMS
x
Square-
root
By Sensor ADC
Digital
lter
Square
RMS
y
Square-
root
Bz Sensor ADC
Digital
lter
Square
RMS
z
Square-
root
RMS
Value
Square-
root
Average
Average
Average
Peak
Value
Peak
detector
Square-
root