8 Functional checks
6. Functional checks
If the final check reveals defects or deviations from
the specified values, ACCUVAC Basic must not be
used again until the faults have been rectified.
We therefore recommend that you always keep a
stock of the following:
Disposable collection bag 21 WM 10732
Aspiration tube with
fingertip 19 WM 10733
6.1 Intervals
To ensure that a properly functioning ACCUVAC Basic is always available, it is essential to observe the fol-
lowing intervals.
Before every use
Perform a functional check (see ”6.2 Perform-
ing the functional check“ on page 8).
After every use
Clean, disinfect and/or sterilize the unit and
its parts (see operating instructions ”5.
Cleaning and disinfecting“);
Perform a functional check (see ”6.2 Perform-
ing the functional check“ on page 8).
Every 3 months
Check the charge level of the power pack 17
by switching on ACCUVAC Basic and read-
ing the voltage indicator 3. If the top LED of the
voltage indicator does not light up, the power
pack should be recharged (see operating in-
structions ”4.4 Charging ACCUVAC Basic“).
At least every 6 months
Perform a functional check (see ”6.2 Perform-
ing the functional check“ on page 8).
Make a visual inspection of the muffler for
clogging. If it is clogged, fit a new muffler (see
”8.14 Changing the muffler“ on page 22).
After all repairs
Clean, disinfect and/or sterilize the unit and
its parts (see operating instructions ”5.
Cleaning and disinfecting“);
Perform a functional check (see ”6.2 Perform-
ing the functional check“ on page 8).
6.2 Performing the functional check
1. Assemble ACCUVAC Basic ready for use.
2. Check that all tubes and the individual parts of
the disposable collection canister 7 are in per-
fect condition. Any damaged or worn parts
must be replaced.
3. Check that all tubes are securely connected
and the disposable collection bag 21 is firmly
installed.
4. Switch on ACCUVAC Basic.
5. Check the charge level on the voltage indica-
tor 3. Always recharge the power pack with-
out delay if the yellow LED on the voltage
indicator lights up (see operating instructions
”4.4 Charging ACCUVAC Basic“).

Other manuals for Weinmann ACCUVAC Basic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann ACCUVAC Basic and is the answer not in the manual?

Weinmann ACCUVAC Basic Specifications

General IconGeneral
BrandWeinmann
ModelACCUVAC Basic
CategoryMedical Equipment
LanguageEnglish

Related product manuals