Safety instructions 13
EN
You may not use it for general operation, as this
would overload and overheat it.
To increase the service life of the battery, it must
not be entirely discharged. Please charge the
battery at the latest when the red LED of the
voltage display lights up.
To counter the risk of a complete discharge, you
must ACCUVAC Basic never store the power
pack when it is discharged. Charge the battery
beforehand according to (see “5.4 Charging
ACCUVAC Basic” on page 26).
Lead batteries discharge even when they are not
in use. This self-discharge increases with increas-
ing temperature and amounts to approx. 20 %
at 20 °C over 3 months without use. Therefore
please comply with the intervals for functional
checks (see “7. Functional check” on page 37).
Do not immerse the bacteria filter in disinfectant
liquid as this adversely affects its hydrophobic
properties.
You must not sterilize the motor unit in an auto-
clave.
You must not immerse the motor unit in a
disinfectant solution.
Do not use any cellphones in the immediate vi-
cinity of ACCUVAC Basic. It is possible to oper-
ate the ACCUVAC Basic without any problems
in the patient area of an ambulance, even if a
cellphone is in use in the cab.