Hygienic preparation 33
EN
3. Lift the disposable canister from its holder.
4. Remove the tube holder plate by pushing it to
the left until it disengages and then pulling it
towards you.
6.2 Cleaning, disinfecting and sterilizing
Hygienic preparation of the ACCUVAC Basic and the
accessories used should be performed as described
in the following table.
Observe the instructions regarding use of disinfectant.
For immersion disinfection we recommend
gigasept
®
FF (new) and terralin
®
protect for wipe dis-
infection.
You are recommended to wear suitable gloves (e.g.
household or disposable gloves) during disinfection
procedures.
Part Cleaning Disinfecting
Rinse in
washing
machine
Sterilization
Disposable
collection canister
Collection canister
In warm water with
a mild detergent
Immerse in dilute
solution
(3)
Rinse at up to
95 °C
Steam
sterilization up
to 121 °C
(4)
Vacuum tube Not permitted
Aspiration tube
with fingertip
Disposable item, re-use not permitted. Use new part for every patient

Other manuals for Weinmann ACCUVAC Basic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann ACCUVAC Basic and is the answer not in the manual?

Weinmann ACCUVAC Basic Specifications

General IconGeneral
BrandWeinmann
ModelACCUVAC Basic
CategoryMedical Equipment
LanguageEnglish

Related product manuals