Functional check 37
7. Functional check
If the functional check reveals defects or deviations
from the specified values, ACCUVAC Basic must not
be used again until the faults have been rectified
(see “8. Troubleshooting” on page 40).
We therefore recommend that you always keep a
stock of the following:
• Disposable collection bag WM 10732
• Aspiration tube with fingertip WM 10733
7.1 Intervals
To ensure that a properly functioning
ACCUVAC Basic is always available, it is essential to
observe the following intervals.
Before every use
• Perform a functional check (see “7.2 Performing
the leak check” on page 38).
After every use
• Clean, disinfect and/or sterilize the unit and its
parts (see “6. Hygienic preparation” on
page 30);
• Clean, disinfect and/or sterilize the vacuum tube
and the disposable collection canister as re-
quired;
• Perform a functional check (see “7.2 Performing
the leak check” on page 38).